Industries Analytical Wine Testing for Winemakers

Analytical Wine Testing for Winemakers

Read time:

Organising and planning the winemaking process is as important as the planting and picking of the fruits. Today’s winemaking process should integrate analytical testing and equipment to test pH, SO2, TA and Brix.
Hanna Instruments have put together a guide of tools and equipment used in winemaking to help you get ready for the season and get accurate and consistent testing results.

How to get ready for wine testing

The following are the essential things you need to do for the successful testing of your wine.

Check your Buffers, Solutions and Reagents

Using fresh products is crucial for successful winemaking. Make sure that buffers, solutions, and reagents are within their stated shelf life. If they have been open for more than six months, they should be replaced.   

Calibration buffers: Always use fresh buffer solution for calibration as this ensures that the buffer being used is not contaminated. Please keep track of how long the solution has been opened and check its expiration date. Sometimes, these solutions may expire before the specified expiration date due to evaporation and influx of atmospheric carbon dioxide.

Hanna Tip: When measuring pH in wine, it is best to calibrate to two points. Most meters will allow for pH 7.01 and pH 4.01 calibration. Ideally, calibration should be done to pH 7.01 and pH 3.00. This allows the calibration to bracket the expected value, which for wine is less than pH 4. Some meters are pre-programmed or allow for a custom calibration point (i.e. pH 3.00).

Electrode Cleaning Solution

electrode Cleaning Solution

Electrodes may become dirty from regular usage and may produce inaccurate readings. A clean electrode is critical for an accurate and stable pH reading. The use of an electrode cleaning solution will eliminate any impurities and residues. Residues can accumulate on the surface of electrodes when immersed in samples during measurement or when stored incorrectly. Additionally, clogging of the junction will result in increased resistance, sluggish and erratic readings, and affect the junction potential. To ensure that the probe is clean and to prevent any clogging of the junction, consider cleaning the bulb and junction of your electrode periodically with an electrode cleaning solution.  

 

lectrode Cleaning Solution

 

Hanna Tip: To remove stains and deposits accumulated on the electrode’s surface, use the Hanna electrode cleaning solution for wine deposits and stains. These solutions are preferred over a general-purpose cleaning solution as they are formulated for a specific purpose.

Electrode Storage Solution

Electrode Storage Solution

The electrode storage solution is specifically formulated for keeping the pH bulb hydrated and maintaining a free-flowing junction. Electrodes must always be kept in storage solution when not in use. The diffusion of the internal electrolyte through the junction will be impeded if the junction is allowed to dry out. The potential of the junction, stability and response time of the electrode will be adversely affected if it is left to dry out. Therefore, electrodes should be stored correctly; otherwise, the life and performance of the electrode will rapidly deteriorate. 

Hanna Tip: 

  • Do not leave the electrode in deionised, purified, or distilled water. This increases the diffusion of the reference electrolyte into the solution and causes water to move into the reference cell by osmosis. Moreover, it can adversely affect the function of your electrode. 
  • Storage solutions are formulated to minimise any concentration gradient between the internal reference and the sample and prevent any organic growth

Electrode Fill Solution

Electrode Fill Solution

As the electrolyte level inside refillable pH or ORP electrode gradually drop over time, it must be topped up with the appropriate electrolyte refill solution to maintain an adequate level. This will ensure a positive head pressure, thus allowing for the electrolyte to flow through the junction into the sample. This is crucial as the ions in the electrolyte electrically connect the meter and electrode with the wine sample being tested. Check the levels of electrodes regularly.  

Hanna Tip: Single junction pH and ORP electrodes use potassium chloride (KCl) solution saturated with silver chloride (AgCl), while double junction electrodes use only KCl. The diagram below will help you to identify what type of electrode you have. 

Single junction pH and ORP electrodes

 

Hanna Buffers and Solutions

  • HI7007L pH 7.01 buffer (500 ml)
  • HI7004L pH 4.01 buffer (500 ml)
  • HI5003 pH 3.00 buffer (500 ml)
  • HI70300L Electrode storage solution (500 mL)
  • HI70635L Cleaning solution for wine deposits (500 mL)
  • HI70636L Cleaning solution for wine stains (500 mL)
  • HI7082 Electrode filling solution for double junction electrodes (4 X 30 mL)
  • HI7071 Electrode filling solution for single-junction electrodes (4 X 30 mL)

   ph solutions

Test and prepare your Electrodes 

Electrodes should be checked for functionality before using them. The overall condition of your electrode depends on offset (pH7.01 mV value) and slope (difference between pH 7.01 mV and 4.01 mV). It is crucial to maintain them, and any deviation from the recommended values will result in erroneous readings.  

The following is the procedure to determine the condition of your electrode and prepare them for your next use. 

  • Use fresh buffers to check the offset and slope of your pH electrode. If the meter you are using has the Good Laboratory Practice (GLP) feature, it will provide the last calibration data. If your meter does not have the GLP feature, you can use any meter with an mV mode to check your pH electrode. 
  • Prepare your electrode by removing its protective/storage cap. If the pH bulb is dehydrated, leave it in a storage solution overnight and remove the fill hole screw cap.
  • Now, it’s time to prepare the calibration buffers. Fill 2 beakers with enough pH buffer solution 7.01 to cover the pH electrode junction. Use the solution in one of the beakers to rinse your pH electrode and the second one for calibration. Repeat this procedure for a pH 3.00 buffer. 
  • Once you have prepared your electrodes and calibration buffers, begin with performing calibration. Start with rinsing your pH electrode in the pH 7.01 rinse beaker and stir gently for 4 to 6 seconds. 
  • Place the electrode in the pH 7.01 calibration beaker and stir gently for 4 to 6 seconds. Then, wait for the reading to stabilise (digits on the LCD stop changing for at least 3 seconds) and confirm the calibration. 
  • Rinse the pH electrode in the pH 3.00 rinse beaker and stir gently for around 8 to 10 seconds. 
  • Finally, place the electrode in the pH 3.00 beaker and stir gently for 4 to 6 seconds. Wait for the pH reading to stabilise and then, confirm the calibration. 

Hanna Tip: 

  • Calibrate the electrode after extended storage, cleaning, and use.
  • If readings are slow to stabilise (more than 30 seconds), the electrode may need to be cleaned, or the electrolyte may need to be replaced.
  • Ensure that fill solution is more than ¾ full
  • Any values outside the +/- 30 mV range indicate that the probe needs to be cleaned and hydrated. If not corrected, then the reference fill solution should be changed in the case of refillable electrodes. If still not corrected, then a new pH electrode should be considered.

Need a new electrode?

Hanna Instruments offer a wide range of pH and ORP electrodes suitable for various applications. For analytical measurements of wine, using an electrode that resists clogging of the junction by the solids in the sample is essential. The following pH and ORP electrodes are recommended for pH/mV meters and titration systems. 

pH

The HI10480 is a digital pH electrode with a built-in temperature sensor. This electrode is specifically designed for measuring pH in wine with a unique outer junction made of PTFE.

Here are some of the key features of the HI10480 pH electrode. 

CPS Sleeve Junction 

CPS™ (Clogging Prevention System) is an innovation in electrode technology. Conventional pH electrodes use ceramic junctions that clog quickly when used in wine. When the junction is clogged, the electrode does not function. CPS™ technology utilises the porousness of ground glass coupled with a PTFE sleeve to prevent clogging of the junction. The ground glass allows proper flow of liquid, while the PTFE sleeve repels dirt. As a result, pH electrodes with CPS™ stay fresh up to 20 times longer than conventional electrodes.

CPS Sleeve Junction 

Double Junction Reference

A double junction electrode has an internal compartment surrounding the reference wire. Silver ions are present in the electrolyte of the internal compartment, which houses the Ag/AgCl reference wire; the electrolyte outside this compartment is silver-free. The double junction design means that virtually no silver from the electrode enters the sample. This design allows measurement in applications where silver ions in the sample are undesirable or silver precipitates on the junction are likely to form.

Double Junction Reference

Refillable probe  

The HI10480 is a refillable probe with a double junction. This fill solution, HI7082 3.5M KCl, does not contain any silver as with a single junction electrode. The absence of silver will prevent any silver precipitate from forming at the junction’s surface and clogging it. Clogging of the junction will result in drifty and erratic readings. 

Refillable probe  

HI10480 Probe by Connector Type (meter dependent) 

The HI10480 pH electrode is a glass body, refillable, sleeve junction with a BNC connector. This pH electrode is ideal for wine and juice due to its high solids content that would clog a standard general-purpose pH electrode.

BNC Connector for HI1048B

BNC connectors are universal sensors. They are compatible with any meter that has a BNC connection. This connector is the original probe that is supplied with the HI84502 acidity mini titrator.  

BNC Connector for HI1048B

BNC and Pin Connectors for HI1048P

The HI1048P is a glass body, refillable, sleeve junction pH electrode with a BNC and pin connector. The pin connector is used to enable the CAL Check function of the HI2221, HI2222, and HI2223 benchtop meters. Probes with BNC connectors and Pin are made specifically for Hanna meters with Pin input for enabling CAL Check.

BNC and Pin Connectors for HI1048P

DIN Connector for HI1048D

The HI1048D is a glass body, refillable, open junction pH electrode with a DIN connector that is supplied with the Portable Wine Must and Grape Juice pH Meter – HI99111. DIN Connectors are proprietary to the meter that uses the connector. The DIN connector allows for multiple wires to be in a single connector. Additional wires are used for temperature measurement and as a power source for amplifiers built into the probe. An amplifier in the probe prevents interference that a temperature signal can have on the pH signal. This threaded series of DIN connectors have a ferrule to attach the probe to the meter.

DIN Connector for HI1048D

3.5 mm Pin Connector for HI10480

This digital connector contains a four-pole design used with digital electrodes. This connection allows digital data transfer from the sensor to the meter and is also meter specific. The probe has an integrated microchip that stores calibration data, electrode type, and serial number, which allows for changing of sensors without having to recalibrate each time. This connector is proprietary and used with the HI2020, HI2030, and HI2040 multiparameter edge meters. The HI10480 can be used with the HI2002 edge dedicated pH meter

3.5 mm Pin Connector for HI10480

ORP Electrodes

ORP Electrodes

ORP Electrodes

Hanna Instruments offer a wide variety of ORP electrodes that are designed for many different applications. 

The HI3148 is a double junction, refillable ORP electrode with a platinum ring for a sensor. The HI3148 has the CPS™ (Clogging Prevention System) technology with the same unique outer junction made of PTFE as the HI1048 pH electrode.

The HI3148B has a BNC connector that can be used with any titrator having a BNC input. The HI3148B is supplied with the HI84500 free and total sulphur dioxide titrator.

Prepare Automatic Titrators and Related Equipment 

Titrators

Titrations can be performed manually; however, automatic titration is the highly preferred technique. Using automatic titrators offers many benefits like improved accuracy and precision, ease of use, and automated calculations. 

The following is the quick guide to prepare automatic titrators and other related equipment.  

Mini Titrator for Measuring SO2 in Wine – HI84500

The HI84500 is a fast, accurate, and affordable automatic mini titrator designed for testing free or total sulphur dioxide (SO2) levels in wine. Based on the Ripper Method, this mini titrator uses an optimised, pre-programmed method of analysis with a powerful algorithm that determines the completion of the titration reaction using a specialised ORP electrode. 

Download quick guide for HI84500

Mini Titrator for Measuring Titratable Acidity in Wine – HI84502

The HI84502 is a simple, fast, and affordable automatic mini titrator designed for testing total acidity levels in wine. Based on an acid-base titration method, this mini titrator uses an optimised pre-programmed method of analysis with a powerful algorithm that determines the completion of the titration reaction using a specialised wine pH electrode. 

Download quick guide for HI84502

Mini Titrator for Measuring Formol Number in Wine – HI84533

The HI84533 is a simple, fast, and affordable automatic mini titrator designed for testing formol numbers in wine.  Based on an acid-base titration method, this mini titrator uses an optimised pre-programmed method of analysis with a powerful algorithm that determines the completion of the titration reaction by using a glass body pH electrode.

Download quick guide for HI84533

 Refractometers

 Refractometers

Consider upgrading your mechanical refractometer to a digital refractometer as it is highly preferred by many due to its ease of use and improved accuracy. Additionally, a digital refractometer will require less maintenance as compared to a mechanical refractometer. The HI96811 and HI96813  are the wine specific refractometers manufactured by Hanna. 

Digital Refractometer for Sugar (% Brix) Analysis in Wine – HI96811

The HI96811 is a rugged, portable digital refractometer designed for sugar measurement in wine.  This refractometer reports sugar content as % Brix and has a high accuracy of ±0.2%. The operation of this refractometer is simplified by having only two buttons: one button is to calibrate using distilled or deionised water, and the other one is to measure. 

Digital Refractometer for % Brix and Potential Alcohol (% V/V) Analysis in Wine – HI96813

The HI96813 is a rugged, portable digital refractometer designed for sugar (% Brix) measurement and potential alcohol measurement (% V/V) in wine. This refractometer features a high accuracy of ±0.2% Brix and 0.2% V/V potential alcohol. All readings are automatically compensated for temperature variations according to the ICUMSA Methods Book standard and displayed within a 1.5 second response time. 

Both refractometers convert the refractive index of wine to % Brix. This conversion is based on the tables found in the ICUMSA Methods Book (International Commission for Uniform Methods of Sugar Analysis) that documents the changes in refractive index with temperature for a percent by weight sucrose solution. Since the majority of sugar in grape juice and in the must is fructose and glucose instead of sucrose, the reading is sometimes referred to as “Apparent Brix”. 

Some of the features of HI96811 and HI96813 are listed below.

  • Reports sugar content as % Brix.
  • Simple operation with only two buttons – one for calibration and the other for measurement.
  • Sample size as small as two metric drops or about 100 μL. 
  • All readings are automatically compensated for temperature variations with a 1.5 second response time
  • The sealed flint glass prism and stainless steel well are easy to clean – just wipe with a soft cloth in preparation for the next sample

Cannot find what you are looking for or need more information about how to prepare for a successful wine testing? Get in touch with a Hanna representative right away. 